Tensile fixation strengths of absorbable meniscal repair devices as a function of hydrolysis time. An in vitro experimental study.

Arnoczky, S P; Lavagnino, M · Am J Sports Med · 2001

biomechanical · Level V

Where this comes from

Abstract

To determine the effect of hydrolysis time on the fixation strength of absorbable meniscal repair devices, adult bovine menisci were repaired with five devices and a suture. The ultimate tensile strength of the repair was then tested in six specimens immediately or after 6, 12, or 24 weeks of incubation at 37 degrees C in a saline solution containing antibiotics, antimycotics, and protease inhibitors. Immediately after implantation the Bionx Meniscus Arrow had a significantly higher failure strength (57.7 +/- 13.8 N) than the Linvatec BioStinger (35.1 +/- 6.7 N), the Innovasive Clearfix screw (34.9 +/- 13 N), the Surgical Dynamics S-D-sorb staple (9.4 +/- 4.6 N), and the Mitek Meniscal Repair System (polydioxanone) (27.2 +/- 6.0 N). However, there was no significant difference between the Bionx Meniscus Arrow and a 2-0 polydioxanone vertical suture (51.6 +/- 2.7 N). The polydioxanone-based implants demonstrated a significant decrease in failure strength at 12 and 24 weeks. Similarly, the Surgical Dynamics S-D-sorb staple lost all fixation strength by 24 weeks. The remaining devices showed no significant loss of failure strength over the 24-week period, suggesting that 24 weeks of hydrolysis does not adversely affect the ultimate holding power of poly L-lactide-based meniscal fixation devices.