Fracture-embolization of duromedics valve prosthesis and microscopic uncommon lesions.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 11308200.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a sudden leaflet fracture of a Duromedics mitral valve 6 years after implantation. The patient had cardiogenic shock and complained of asthenia, orthopnea, and tachycardia. Transesophageal echocardiography showed the lack of one leaflet of the prosthesis and regurgitation. An emergency mitral replacement was successfully performed. Angiographic computed tomography scan localized the sequestrum that embolized the common iliac arteries. Examination of the deficient prosthesis showed multiple lesions and, in particular, a subsurface lesion that may be characteristic of carbon pyrolytic valves.
Medical subject headings
- Bioprosthesis
- Heart Valve Prosthesis
- Prosthesis Failure
- Prosthesis-Related Infections
- Streptococcal Infections