Flow effect on lesion formation in RF cardiac catheter ablation.

Cao, H; Vorperian, V R; Tungjitkusolmun, S; Tsai, J Z; Haemmerich, D; Choy, Y B; Webster, J G · IEEE Trans Biomed Eng · 2001

basic_science · Level V

Where this comes from

Abstract

This study investigated the flow effect on the lesion formation during radio-frequency cardiac catheter ablation in temperature-controlled mode. The blood flow in heart chambers carries heat away from the endocardium by convection. This cooling effect requires more power from the ablation generator and causes a larger lesion. We set up a flow system to simulate the flow inside the heart chamber. We performed in vitro ablation on bovine myocardium with three different flow rates (0 L/min, 1 L/min and 3 L/min) and two target temperatures (60 degrees C and 80 degrees C). During ablation, we also recorded the temperatures inside the myocardium with a three-thermocouple temperature probe. The results show that lesion dimensions (maximum depth, maximum width and lesion volume) are larger in high flow rates (p<0.01). Also, the temperature recordings show that the tissue temperature rises faster and reaches a higher temperature under higher flow rate.

Medical subject headings