Clinical thermometry, using the 27 MHz multi-electrode current-source interstitial hyperthermia system in brain tumours.

Kaatee, R S; Nowak, P C; van der Zee, J; de Bree, J; Kanis, B P; Crezee, H; Levendag, P C; Visser, A G · Radiother Oncol · 2001

basic_science · Level V

Where this comes from

Abstract

In interstitial hyperthermia, temperature measurements are mainly performed inside heating applicators, and therefore, give the maximum temperatures of a rather heterogeneous temperature distribution. The problem of how to estimate lesion temperatures using the multi-electrode current-source interstitial hyperthermia (MECS-IHT) system in the brain was studied. Temperatures were measured within the electrodes and in an extra catheter at the edge of a 4 x 4 x 4.5 cm(3) glioblastoma multiforme resection cavity. From the temperature decays during a power-off period, information was obtained about local maximum and minimum tissue temperatures. The significance of these data was examined through model calculations. Maximum tissue temperatures could be estimated roughly by switching off all electrodes for about 5 s. Model calculations showed that the minimum tissue temperatures near a certain afterloading catheter correspond well with the temperature of the applicator inside, about 1 min after this applicator was switched off. Although the electrode temperatures read during heating are not suitable to assess the temperature distribution, it is feasible to heat the brain adequately using the MECS-IHT system with extra sensors outside the electrodes and/or application of decay methods.