Processing of HA-coated Ti-6Al-4V by a ceramic slurry approach: an in vitro study.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11336294.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Hydroxyapatite-coated titanium alloy composite powders (Ti-6Al-4V/HA) was produced by a ceramic slurry approach. The aim was to evaluate the stability of the coating when subjected to a physiological medium simulated body fluid (SBF). Three consolidation conditions (700 degrees C for 5 h, 700 degrees C for 8 h and 700 degrees C for 11 h) were used in the production of the Ti-6Al-4V/HA composite powders. Results showed that biodissolution followed by apatite precipitation had taken place after soaking in SBF. In addition, it was found that consolidation at 700 degrees C for 5 h resulted in a weak mechanical locking of calcium phosphate on the Ti-6Al-4V surfaces; and the formation of small crystallites, which would increase the dissolution rate.
Medical subject headings
- Ceramics
- Durapatite
- Titanium