Pitfalls associated with fixation of osteochondritis dissecans fragments using bioabsorbable screws.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11337726.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate 2 cases in which bioabsorbable screw fixation for an osteochondritis dissecans lesion of the femoral condyle resulted in complications necessitating the need for secondary surgery. We reviewed the case history of these patients and described the circumstances under which the bioabsorbable screws were used, the events leading to the need for secondary surgery, and the ultimate outcome. In the 2 cases presented, these implants were found to retain their mechanical stiffness for many months. This resulted in articular damage in 1 case after the treated lesion failed to heal. In the second case, screw breakage 8 months after implantation resulted in it becoming a loose body, which required removal during a second arthroscopic procedure. We conclude that these implants retain their mechanical properties for many months and cannot be relied on to degrade quickly. If a treated lesion fails to heal, these implants can cause mechanical problems due to their retained structural properties.
Medical subject headings
- Absorbable Implants
- Bone Screws
- Femur
- Occupational Diseases
- Osteochondritis Dissecans
Anatomy
- femur
- knee