Optimising miller cuff dimensions: the influence of geometry on anastomotic flow patterns.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11352685.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
since cuff dimensions are variable, we studied the influence of cuff geometry on flow mechanics, in an attempt to identify the optimum configuration. bench studies involved the manufacture of anatomically accurate models of varying cuff dimensions, perfused in a specifically designed flow rig, simulating physiological conditions. Flow visualisation studies incorporating laser illumination of tracer particles enabled accurate analysis of flow patterns. the vortex created within the proximal cuff of each model during the deceleration phase of the cardiac cycle was strongly influenced by the aspect ratio (AR=cuff height:length). The standard and high cuffs (AR=1.63 and 1.18, respectively) demonstrated cohesive vortices and stable flow patterns. Low and long cuffs (AR=2.6 and 2.25, respectively) created more complex vortices with large areas of flow separation and low velocities. aspect ratio has an important influence on flow within the distal anastomosis, with cuff dimensions of 13 mm long and 8-11 mm high (standard and high cuffs) creating beneficial flow patterns anticipated to optimise wall shear stress and inhibit myointimal hyperplasia.
Medical subject headings
- Anastomosis, Surgical
- Models, Cardiovascular
- Rheology