Evaluation of in vivo dose measurements for patients undergoing electron boost treatments.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 11369070.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study evaluated p-type silicon diodes for use in in vivo dosimetry in clinical electron beams. A calibrated p-type silicon diode detector was used to measure the dose received by the patient in the centre of the field. Readings were corrected for energy, temperature and stand-off of the electron applicator from the patient surface. The mean difference between measured and prescribed dose was 1.04% (95% CI 0.72 to 1.36 %).