Electrocautery-induced tachycardia in a rate-responsive pacemaker.

Wong, D T; Middleton, W · Anesthesiology · 2001

case_report · Level V

Where this comes from

Abstract

Electric interference by cautery on demand pacemakers can cause inhibition and loss of pacing. We report a case in which electrocautery induced a pacemaker in ventricle-paced, ventricle-sensed, inhibited, rate-responsive (VVIR) mode to pace at a programmed maximum rate of 130 pulses/min.