The development of solid nails has not eliminated the risk of occasional nail failure. It has been suggested that subsequent intra-medullary removal of solid nail fragment may be virtually impossible and thus major surgery is required for their revision. We report two cases of hardware failure of a solid intra-medullary nail (Synthes solid femoral and tibial nail) and describe their successful intra-medullary removal using the Synthes extraction kit.