Diagnostic laryngeal electromyography: The Wake Forest experience 1995-1999.

Koufman, J A; Postma, G N; Whang, C S; Rees, C J; Amin, M R; Belafsky, P C; Johnson, P E; Connolly, K M et al. · Otolaryngol Head Neck Surg · 2001

retrospective_cohort · Level III

Where this comes from

Abstract

Laryngeal electromyography (LEMG) is a valuable diagnostic/prognostic test for patients with suspected laryngeal neuromuscular disorders. To report our experience with diagnostic LEMG at the Center for Voice Disorders of Wake Forest University and to evaluate the impact of LEMG on clinical management. Retrospective chart review of 415 patients who underwent diagnostic LEMG over a 5-year period (1995-1999). Of 415 studies, 83% (346 of 415) were abnormal, indicating a neuropathic process. LEMG results altered the diagnostic evaluation (eg, the type of radiographic imaging) in 11% (46 of 415) of the patients. Unexpected LEMG findings (eg, contralateral neuropathy) were found in 26% (107 of 415) of the patients, and LEMG results differentiated vocal fold paralysis from fixation in 12% (49 of 415). Finally, LEMG results altered the clinical management (eg, changed the timing and/or type of surgical procedure) in 40% (166 of 415) of the patients. LEMG is a valuable diagnostic test that aids the clinician in the diagnosis and management of laryngeal neuromuscular disorders.

Medical subject headings