An ultralight biotelemetry backpack for recording EMG signals in moths.

Mohseni, P; Nagarajan, K; Ziaie, B; Najafi, K; Crary, S B · IEEE Trans Biomed Eng · 2001

basic_science · Level V

Where this comes from

Abstract

A two-channel FM biopotential recording system fabricated on a foldable, lightweight, polyimide substrate is presented. Each channel consists of a biopotential amplifier followed by a Colpitts oscillator with operating frequency tunable in the 88-108 MHz commercial FM band. The overall system measures 10 mm X 10 mm X 3 mm, weighs 0.74 g, uses two 1.5-V batteries, dissipates about 2 mW, and has a transmission range of 2 m. Using this system, electromyogram signals have been recorded from the dorsal ventral muscle and the dorsal longitudinal muscle of a giant sphinx moth (manduca sexta).

Medical subject headings