Self-expandable chitosan stent: design and preparation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11396892.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Stents are largely used in surgical procedures to relieve pathological obstructions. The purpose of the present study was to design and prepare a biocompatible stent with a self-expandable mechanism. Thin films were prepared from deacetylated chitosan (4% w/v) dissolved in acetic acid solution (2% v/v). The chitosan films were tested by a calibrated tensiometer to measure the Young's module (E). The films were used to manufacture stents by pulling and winding them around a cylindrical rod in a helical fashion. Thirteen stents (diameter = 0.5 +/- 0.05 mm, length approximately 4 mm) were inserted into the vas deferens of wistar rats. Upon stent insertion, the vasal anastomosis was achieved with a laser-soldering technique. The animals were sacrificied 8 weeks later. The stress test showed that the chitosan film was elastic (maximum strain = 105% +/- 6%, E = 0.7655 +/- 0.0288 Mpa). The stents self-expanded by releasing their elastic energy. All the stents but one remained open inside the vasa despite high incidence of sperm granuloma. A biocompatible and self-expandable stent with a helical design is proposed.
Medical subject headings
- Biocompatible Materials
- Chitin
- Equipment Design
- Stents