Head penetration into Hylamer acetabular liner sterilized by gamma irradiation in air and in a nitrogen atmosphere.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 11402409.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We reviewed 25 consecutive primary cementless total hip arthroplasties with Hylamer acetabular liners (Hylamer group) and 12 with conventional ultra-high molecular weight polyethylene (Enduron group). Two-dimensional penetration of the femoral head into the liner was determined from anteroposterior radiographs of the pelvis. Head penetration rate was 0.37 mm/y in the Hylamer group sterilized by gamma irradiation in air (n = 6; mean length of follow-up, 3 years), 0.21 mm/y in the Hylamer group sterilized by gamma irradiation in a nitrogen atmosphere (n = 19; mean length of follow-up, 2.7 years), and 0.11 mm/y in the Enduron group (n = 12; mean length of follow-up, 3.9 years). Osteolysis was identified in 6 of the 25 hips with Hylamer liners and 1 of the 12 hips with conventional liners. There was a positive linear correlation between period from production to operation and head penetration rate with Hylamer liner sterilized by gamma irradiation in air and no correlation in a nitrogen atmosphere. Rapid oxidation by irradiation in air might not be the main cause of high rate of wear in Hylamer liners.
Medical subject headings
- Gamma Rays
- Hip Prosthesis
Anatomy
- hip
- pelvis