Comments on the filtered backprojection algorithm, range conditions, and the pseudoinverse solution.
editorial · Level V
Where this comes from
- Record sourced from PubMed, PMID 11437114.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The filtered backprojection (FBP) algorithm is widely used in computed tomography for inverting the two-dimensional Radon transform. In this paper, we analyze the processing of an inconsistent data function by the FBP algorithm (in its continuous form). Specifically, we demonstrate that an image reconstructed using the FBP algorithm can be represented as the sum of a pseudoinverse solution and a residual image generated from an inconsistent component of the measured data. This reveals that, when the original data function is in the range of the Radon transform, the image reconstructed using the FBP algorithm corresponds to the pseudoinverse solution. When the data function is inconsistent, we demonstrate that the FBP algorithm makes use of a nonorthogonal projection of the data function to the range of the Radon transform.
Medical subject headings
- Algorithms
- Image Processing, Computer-Assisted
- Tomography, X-Ray Computed