Dipstick measurements of urine specific gravity are unreliable.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 11466191.
- Also identified by PMC identifier 1718890.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the reliability of dipstick measurements of urine specific gravity (U-SG). Fresh urine specimens were tested for urine pH and osmolality (U-pH, U-Osm) by a pH meter and an osmometer, and for U-SG by three different methods (refractometry, automatic readout of a dipstick (Clinitek-50), and (visual) change of colour of the dipstick). The correlations between the visual U-SG dipstick measurements and U-SG determined by a refractometer and the comparison of Clinitek((R))-50 dipstick U-SG measurements with U-Osm were less than optimal, showing very wide scatter of values. Only the U-SG refractometer values and U-Osm had a good linear correlation. The tested dipstick was unreliable for the bedside determination of U-SG, even after correction for U-pH, as recommended by the manufacturer. Among the bedside determinations, only refractometry gives reliable U-SG results. Dipstick U-SG measurements should be abandoned.