Evaluation of image quality and dose in renal colic: comparison of different spiral-CT protocols.
Level III
Where this comes from
- Record sourced from PubMed, PMID 11471601.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to test different technical spiral-CT parameters to obtain optimal image quality with reduced X-ray dose. Images were acquired with a spiral-CT system Philips Tomoscan AVE1, using 250 mA, 120 kV, and 1-s rotational time. Three protocols were tested: protocol A with 5-mm thickness, pitch 1.6, slice reconstruction every 2.5 mm; protocol B with 3-mm thickness, pitch 1.6, slice reconstruction every 1.5 mm; and protocol C with 3-mm thickness, pitch 2, slice reconstruction every 1.5 mm. Two phantoms were employed to evaluate the image quality. Axial images were acquired, then sagittal and coronal images were reconstructed. Finally, the absorbed X-ray dose for each protocol was measured. Regarding image quality, 5-mm-thick images (protocol A) showed greater spatial resolution and lower noise compared with 3-mm-thick images (protocols B and C) on the axial plane; 3-mm reconstructed sagittal and coronal images (protocols B and C) showed an improved image quality compared with 5-mm reformatted images (protocol A). Concerning X-ray dose, the mean dose was: protocol A 19.6 +/- 0.8 mGy; protocol B 14.4 +/- 0.6 mGy; protocol C 12.5 +/- 1.0 mGy. Our study supports the use of thin slices (3 mm) combined with pitch of 1.6 or 2 in renal colic for X-ray dose reduction to the patient and good image quality.
Medical subject headings
- Colic
- Image Enhancement
- Kidney Diseases
- Tomography, X-Ray Computed