Image quality vs. radiation dose for a flat-panel amorphous silicon detector: a phantom study.
other
Where this comes from
- Record sourced from PubMed, PMID 11511892.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of this study was to investigate the image quality for a flat-panel amorphous silicon detector at various radiation dose settings and to compare the results with storage phosphor plates and a screen-film system. A CDRAD 2.0 contrast-detail phantom was imaged with a flat-panel detector (Philips Medical Systems, Eindhoven, The Netherlands) at three different dose levels with settings for intravenous urography. The same phantom was imaged with storage phosphor plates at a simulated system speed of 200 and a screen-film system with a system speed of 160. Entrance surface doses were recorded for all images. At each setting, three images were read by four independent observers. The flat-panel detector had equal image quality at less than half the radiation dose compared with storage phosphor plates. The difference was even larger when compared with film with the flat-panel detector having equal image quality at approximately one-fifth the dose. The flat-panel detector has a very favourable combination of image quality vs radiation dose compared with storage phosphor plates and screen film.