A reevaluation and discussion on the threshold limit for hemolysis in a turbulent shear flow.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11522317.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
One major factor of red blood cell damage induced by artificial heart valves is the magnitude of turbulent shear stresses in the flow field. An often-cited threshold for hemolysis is 400 N/m(2) (Sallam and Hwang, Biorheology 21 (1984) 783). This value, however, was measured with a one-component laser Doppler anemometer and was not calculated from the major principal Reynolds shear stress, therefore underestimating the threshold. This study performed flow field measurements under similar conditions, using a two-component laser Doppler anemometer and determining major principal Reynolds shear stress, to reevaluate the hemolytic threshold as 800 N/m(2) with an exposure time of 1 ms. The Kolmogorov length scales, approximately 9 microm, are capable of inflicting direct damage to the red blood cells. These results will serve as a more accurate reference in future heart valve design and testing.
Medical subject headings
- Heart Valve Prosthesis
- Hemolysis