Right atrial septal electrode for reducing the atrial defibrillation threshold.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11524403.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The atrial defibrillation threshold (ADFT) energy of the standard lead configuration, right atrial appendage (RAA) to coronary sinus (CS), was reduced by >50% with the addition of a third electrode traversing the atrial septum in a previous study. This study determined whether the ADFT would be lowered by a more clinically practical third electrode placed in the right atrium along the atrial septum (RSP). Sustained atrial fibrillation was induced in 8 closed-chest sheep with burst pacing and maintained with pericardial infusion of acetyl-beta-methylcholine chloride. A custom-made, dual-defibrillation catheter was placed with electrodes in the lateral RA, CS, and RSP. A separate defibrillation catheter was also placed in the RAA. ADFT characteristics of RAA-->CS and 6 other single- or sequential-shock configurations were determined in random order by using biphasic, truncated-exponential waveforms in a multiple-reversal protocol. The delivered-energy, peak-voltage, and peak-current ADFTs for the sequential-shock configuration CS-->RSP/RA-->RSP (0.53+/-0.31 J, 86+/-22 V, and 1.6+/-0.6 A, respectively) were significantly lower than those of RAA-->CS (1.14+/-0.64 J, 157+/-34 V, and 2.5+/-1.1 A, respectively). The ADFT characteristics of RAA-->CS and RA-->CS were not significantly different, nor were those of CS-->RSP/RA-->RSP and CS-->RSP/RAA--> RSP. The ADFT of the standard RAA-->CS configuration may be markedly reduced with an additional electrode situated at the RSP.
Medical subject headings
- Atrial Fibrillation
- Electrodes, Implanted
- Heart Atria
- Heart Septum