Evaluation of a new percutaneous dilatational tracheostomy set apparatus.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11531672.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Percutaneous tracheostomy is a well established technique used primarily to assist weaning from mechanical ventilation on many intensive care units. We report our experiences of a total of 36 procedures performed with the new Blue Rhino Percutaneous Tracheostomy Introducer Set developed by Ciaglia. The technique was successful in all cases and was simpler and quicker to perform than with the earlier Ciaglia percutaneous tracheostomy set. Difficulties were encountered when using Shiley tracheostomy tubes. Significant complications included one posterior wall tear and one tracheal cartilage ring fracture.
Medical subject headings
- Critical Care
- Tracheostomy