The strength of a cement acetabular locking mechanism.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11547373.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To minimize the morbidity associated with revision of a well-fixed acetabular shell, some surgeons advocate cementing a polyethylene liner into the shell when a liner exchange is not possible. So far the strength of this cement locking mechanism has been measured on only a few specimens. The purpose of this study is to correct that oversight by measuring the push-out and lever-out strengths of an acetabular polyethylene liner cemented into a porous-coated shell. We cemented acetabular liners into porous-coated acetabular shells using first-generation cementing techniques. Each of 30 specimens was tested for its push-out or lever-out strength. The results suggest that the strength of a cemented polyethylene liner locking mechanism is similar to that of an uncemented acetabular shell.