Non-cycloplegic screening for amblyopia via refractive findings with the Nikon Retinomax hand held autorefractor in 3 year old kindergarten children.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11567961.
- Also identified by PMC identifier 1723736.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess non-cycloplegic screening for amblyopia with the hand held Nikon Retinomax autorefractor in 3 year old kindergarten children. 427 three year old children were examined in kindergarten with the Retinomax without cycloplegia. A gold standard was established in all children by two orthoptic examinations in kindergarten. If there were missing, abnormal, or inconsistent findings, children were referred for ophthalmological examination. If, by the ophthalmological examination, a new case of amblyopia requiring treatment was diagnosed, the gold standard was set "positive." In 404 children the gold standard was obtained. 10 children (2.5%) had a "positive" gold standard of unknown and untreated amblyopia. Screening sensitivity was 0.80, specificity 0.58, accuracy 0.58, and the likelihood ratio 1.89. Non-cycloplegic refractive screening with the Retinomax led to many false positive referrals due to instrument myopia and "inconclusive" results. Hence specificity, accuracy, and the likelihood ratio were too low to conduct screening effectively.
Medical subject headings
- Amblyopia
- Mass Screening
- Mydriatics
- Ophthalmology
- Refraction, Ocular