Compound A production from sevoflurane is not less when KOH-free absorbent is used in a closed-circuit lung model system.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11573522.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In an in vitro study, less compound A was formed when a KOH-free carbon dioxide absorbent was used. To confirm this observation we used a lung model in which carbon dioxide was fed in at 160 ml min(-1) and sampling gas was taken out for analysis at 200 ml min(-1); ventilation aimed for a PE'CO2 of 5.4 kPa. The soda lime canister temperatures in the inflow and outflow ports (Tin and Tout) were recorded. In six runs of 240 min each, a standard soda lime, Sodasorb (Grace, Epernon, France) was used and in eight runs KOH-free Sofnolime (Molecular Products, Thaxted, UK) was used. Liquid sevoflurane was injected using a syringe pump to obtain 2.1% E'. Compound A was measured by capillary gas chromatography combined with mass spectrometry. Median (range) compound Ainsp increased to a maximum of 22.7 (7.9) ppm for Sodasorb and 33.1 (20) for Sofnolime at 60 min and decreased thereafter; the difference between groups was significant (P<0.05) at each time of analysis up to 240 min. The canister temperatures were similar in both groups and increased to approximately 40 degrees C at 240 min. Contrary to expectation, compound A concentrations were greater with the KOH-free absorbent despite similar canister temperatures with both absorbents.
Medical subject headings
- Anesthesia, Closed-Circuit
- Anesthetics, Inhalation
- Calcium Compounds
- Ethers
- Hydrocarbons, Fluorinated
- Methyl Ethers
- Oxides
- Sodium Hydroxide