Mechanical properties of artificial tracheas composed of a mesh cylinder and a spiral stent.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11603578.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Much work has been done on the materials used for mesh-type artificial tracheas, but a precise mechanical evaluation of these structures has not yet been performed. In the present study, we determined the mechanical properties of typical mesh-type artificial tracheas and compared them with those of native trachea. Four types of artificial trachea were made and used for the mechanical tests. The basic frame of all the specimens was composed of a mesh cylinder and a spiral stent. The specimen whose mesh was sealed with collagen sponge showed almost the same behavior in the force-strain curve under compression, suggesting that collagen sealing has little effect on mechanical properties. Agreement between measured and estimated mechanical properties was good, especially in the low strain region, suggesting that artificial tracheas can be designed in terms of mechanical properties by mainly considering the basic frame structure.
Medical subject headings
- Artificial Organs
- Materials Testing
- Stents
- Trachea