Early loosening of the stemmed McMinn cup.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11607903.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The stemmed McMinn cup was used in 26 acetabular reconstructions (median age, 59 [range, 33-85 years]; 21 women) at 5 orthopaedic centers in Sweden between 1995 and 1998. Of the patients, 24 (92%) had severe bone defects (Gustilo-Pasternak grade 3 and 4). At a clinical and radiographic follow-up of median 3 years (range, 1-5 years), 4 (17%) cups were rerevised at 1 (2 cups), 2, and 3 years owing to symptomatic aseptic loosening. Another 8 cups (40%) were radiographically definitely loose and 1 (5%) was probably so. With an overall mechanical failure rate (rerevised plus definitely loose devices) of 43.8% at a minimum 3-year follow-up, the stemmed McMinn cup does not seem to be a reliable solution in acetabular reconstruction.