Rigid registration of 3-D ultrasound with MR images: a new approach combining intensity and gradient information.

Roche, A; Pennec, X; Malandain, G; Ayache, N · IEEE Trans Med Imaging · 2001

other · Level V

Where this comes from

Abstract

We present a new image-based technique to rigidly register intraoperative three-dimensional ultrasound (US) with preoperative magnetic resonance (MR) images. Automatic registration is achieved by maximization of a similarity measure which generalizes the correlation ratio, and whose novelty is to incorporate multivariate information from the MR data (intensity and gradient). In addition, the similarity measure is built upon a robust intensity-based distance measure, which makes it possible to handle a variety of US artifacts. A cross-validation study has been carried out using a number of phantom and clinical data. This indicates that the method is quite robust and that the worst registration errors are of the order of the MR image resolution.

Medical subject headings