Spatial resolution of a stacked radiochromic film dosimeter.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11690689.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The spatial resolution of stacked radiochromic film dosimeters, which have increased sensitivity from a single layer radiochromic film detector, has been studied. A 5-layer film which can easily be constructed provided a 4.3-times increase in sensitivity over a single layer film at 670 nm readout wavelength which meant that doses as low as 0.6 Gy could be measured with an accuracy of +/-4% with the stacked dosimeter. The spatial resolution was tested by comparison of the 80%/20% penumbral widths of a 5 x 5 cm 6 MV X-ray field. The MD-55-2 film measured the penumbral width as 3.0 mm whereas the 5-layer stack dosimeter measured the same penumbra as 3.2 mm. The stack dosimeter can provide useful in vivo information such as the position of a diverging beam edge for treatments around critical structures such as eyes during the first fraction of treatment.