Pilot study of the accuracy of bedside glucometry in the intensive care unit.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 11700424.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the accuracy of bedside glucometry among critically ill patients. Prospective audit, with sequential specimen collection. Thirty-two-bed cardiovascular, neurosurgical, and medical-surgical intensive care unit in a single Canadian center. Ten critically ill adults, who were sequentially followed during their intensive care unit stay. Eight had diabetes mellitus, and three were in shock. Repeat arterial blood samples were obtained by the attending nurse, who withdrew a portion of the sample and performed reflectance glucometry at the bedside. The remainder was immediately sent in a vacuum-sealed plasma separation tube to the hospital laboratory, and analyzed using a conventional plasma glucose analyzer by a laboratory technologist. Sequential samples were taken at intervals of at least 12 hrs of one other. A total of 105 arterial glucose pairs were obtained. There was a significant correlation between the laboratory and glucometry determined glucose concentrations (intraclass correlation coefficient = 0.86, p <.0001). The overall average laboratory-glucometry glucose difference was -0.04 mmol/L (95% confidence interval [CI] -2.3-2.2 mmol/L). Five out of 105 values (4.8%) lay beyond these confidence bounds. The square of the Pearson correlation coefficient (r(2)) between the mean glucose level and the laboratory-glucometry glucose difference was not significant (0.01, 95% CI 0.005-0.04; p =.22), suggesting the absence of any trend between rising glucose concentration and the laboratory-glucometry difference. Bedside glucose testing of arterial whole blood samples may be an accurate alternative to laboratory plasma glucose measurement among critically ill adults, within approximately 2.3 mmol/L of certainty. Because previous studies have suggested that this bedside technique may be prone to a moderate degree of error among patients in shock as well as those with an abnormal blood pH or hematocrit, larger studies are needed to confirm our findings.
Medical subject headings
- Blood Glucose
- Intensive Care Units
- Point-of-Care Systems