A study on beams passing through hip prosthesis for pelvic radiation treatment.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11704342.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To study the dose distributions at the interface due to the presence of a metal implant; to show the dose distributions in combined fields in the presence of hip prostheses; and to demonstrate the capabilities and limitations of a conventional system. Perturbations in the dose distribution caused by a hip prosthesis can result in unacceptable dose inhomogeneities within the target volume and in regions where tissues interface with implant. The Monte Carlo technique and a conventional treatment planning system are used to calculate the dose distributions. Dose increases of 15% in tissue are seen at the interface between metal implant and tissue. Dose reductions of 5-25% or 10-45% are observed in the shadow of the hip prosthesis made of 0.5-3-cm-thick titanium or steel alloy respectively. We compared predicted dose distribution between the Monte Carlo simulation and a commercial treatment planning system (CADPLAN). We found that CADPLAN underestimated the attenuation of hip prostheses. This has led to overestimation of the target dose by 14% for a typical four-field box technique. An acceptable dose distribution can be achieved with a proper lateral beam weighting and compensation using an eight-field technique. The beam compensation may be applied to achieve an adequate target dose.
Medical subject headings
- Hip Prosthesis
- Prostatic Neoplasms