Relationship between gravimetric wear and particle generation in hip simulators: conventional compared with cross-linked polyethylene.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11712832.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Hip-simulator studies have shown reduced gravimetric wear rates for inert-gas gamma-irradiated ultra-high molecular weight polyethylene when compared with conventional ethylene-oxide-sterilized ultra-high molecular weight polyethylene. Analysis shows a greater number of particles generated from inert-gas gamma-irradiated ultra-high molecular weight polyethylene. This study was undertaken to examine particle-generation rates of polyethylene with different levels of cross-linking and to correlate them with gravimetric wear data. Particle-generation rates did not correlate with gravimetric wear rates. Particle analysis should be performed to predict the in vivo behavior of bearing surface materials. Cross-linked ultra-high molecular weight polyethylene subjected to 10 Mrad (100,000 Gy) of gamma irradiation generated significantly fewer particles than ethylene-oxide-sterilized ultra-high molecular weight polyethylene; it also demonstrated a 96% reduction in the volume of particles.
Medical subject headings
- Hip Prosthesis
- Polyethylenes
- Prosthesis Failure