Linear and volumetric wear of tibial inserts in posterior cruciate-retaining knee arthroplasties.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11716374.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Linear and volumetric wear was measured in 33 tibial polyethylene inserts from three different cruciate-retaining knee systems retrieved at the time of revision surgery. Wear patterns also were evaluated and classified. Eccentric and asymmetric wear patterns were seen in 78% of inserts with flat articulating geometry versus 12% in inserts with curved anteroposterior geometry. The mean linear wear rate was .35 mm/year (range, .05-1.68 mm/year) and the mean volumetric wear rate was 794 mm3/year (range, 24-4088 mm3/year). Linear and volumetric wear rates showed a negative correlation with the length of implantation. Linear wear rates also showed a negative correlation with patient weight.
Medical subject headings
- Knee Prosthesis
- Polyethylenes
Anatomy
- knee
- tibia