Triple taper polished cemented stem in total hip arthroplasty: rationale for the design, surgical technique, and 7 years of clinical experience.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 11742449.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Successful long-term clinical results with the Charnley low-friction torque arthroplasty have identified proximal femoral strain shielding as a long-term problem to be addressed. The problem has arisen because of the load transfer by a distally supported stem that is no longer subject to fracture. To overcome this problem and improve proximal load transfer to the femur, a continuous triple-tapered stem--the C stem--was designed. This article describes the first 500 primary hip arthroplasties at a mean follow-up of 3 years, 5 months (range, 1-7 years). There have been no revisions for aseptic stem loosening, and no stem is considered to be at risk for loosening. In 20% of cases, there was subjective radiologic improvement of the bone-cement interface.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- hip