Comparison of femoral head penetration rates between cementless acetabular components with 22-mm and 28-mm heads.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 11742461.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Between April 1993 and July 1994, the senior author (R.C.J.) performed 151 consecutive primary total hip arthroplasties using a Harris-Galante II acetabular component and a polished Iowa femoral component. In 105 hips, 22-mm head components were used, and in 46 hips, 28-mm head components were used. Femoral head penetration into the acetabular shell was measured using digital edge detection techniques. The group average penetration for the initial 2-year bedding-in period was 0.35 mm/y for the 22-mm and 0.31 mm/y for the 28-mm head components. The long-term rate of penetration into the shell was 0.11 mm/y for the 22-mm heads and 0.17 mm/y for the 28-mm heads, a significant difference (P=.029). The dislocation rate was significantly higher, however, with the 22-mm heads (P=.001). The 22-mm components showed significantly less wear but at the expense of an increase in the prevalence of dislocation.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- femur
- hip