Point-of-care testing for prothrombin time, but not activated partial thromboplastin time, correlates with laboratory methods in patients receiving aprotinin or epsilon-aminocaproic acid while undergoing cardiac surgery.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 11789734.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Point-of-care testing (POCT) of coagulation parameters can help optimize transfusion practice in cardiac surgery. Antifibrinolytic agents may interfere with the laboratory and/or POCT coagulation assays. This randomized controlled study compared coagulation parameters obtained from a whole blood POCT coagulation device with a typical laboratory instrument in cardiac surgery patients receiving aprotinin, epsilon-aminocaproic acid, or normal saline before undergoing cardiopulmonary bypass. Aliquots of arterial blood samples from 42 patients were collected perioperatively, and their prothrombin times (PTs) and activated partial thromboplastin times (aPTTs) were measured by POCT and laboratory instrumentation. Linear regression and error analyses were used for the method comparison. For PT, the POCT device compared favorably with the laboratory method. For aPTT, the POCT device did not compare well with the laboratory method. Treatment with antifibrinolytic agents does not interfere with determination of PT.
Medical subject headings
- Aminocaproic Acid
- Antifibrinolytic Agents
- Aprotinin
- Hemostasis, Surgical
- Partial Thromboplastin Time
- Point-of-Care Systems
- Prothrombin Time