Dosimetric verification of modulated photon fields by means of compensators for a kernel model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 11830316.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The approach in treatment planning of applying beam quality correction factors to model compensator-induced depth-hardening effects is investigated and the present work comprises a dosimetric verification of the model for a common compensator material. Lead sheet modulators for four different phantom shapes were designed using a treatment planning system based on the model. The modulators were designed to yield homogeneous dose in a plane. The calculated modulation created by the lead sheets was re-imported into the treatment planning system and applied to a water phantom geometry for verification purposes. Comparing measurements, a total of 31 different geometries were measured, with calculations in this geometry showing good agreement for depth doses, dose profiles and output data with a maximum deviation of 4% except locally in the penumbra region and close to the edges of the cut lead sheets.
Medical subject headings
- Photons
- Radiotherapy Dosage
- Radiotherapy Planning, Computer-Assisted