Cardiac function: MR evaluation in one breath hold with real-time true fast imaging with steady-state precession.
case_control · Level III
Where this comes from
- Record sourced from PubMed, PMID 11867810.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In 12 healthy volunteers and eight patients with cardiac disease, cine magnetic resonance (MR) imaging in the heart was performed with real-time true fast imaging with steady-state precession (FISP), which permitted evaluation of the entire left ventricle in one breath hold (91 msec per frame, 13 frames per section position, nine short-axis section positions per breath hold). Contrast-to-noise ratios (CNRs) and left ventricular mass and function measurements with this technique were compared in all subjects with single-section true FISP imaging and, in the volunteers only, with segmented fast low-angle shot (FLASH) MR imaging. Myocardium-to-blood CNR was significantly higher for both true FISP sequences compared with the FLASH sequence. Measurements of resting left ventricular function with real-time true FISP imaging were comparable with those derived from a series of separate breath-hold single-section true FISP acquisitions.