Prosthetic valve malfunction masked by intraoperative pressure measurements.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 11916784.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We describe a case in which intraoperative echocardiography recorded an abnormally high pressure gradient across a newly implanted mechanical heart valve. However, inserting pressure-transducing needles on each side of this prosthesis did not confirm the echocardiographic findings. The prosthesis was later confirmed to be malfunctioning and was replaced.