Evaluation of ventilators used during transport of ICU patients -- a bench study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 11967599.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate portable ventilators. Bench study. Five portable ventilators used for transporting ICU patients [Osiris 1, (ventilator a), Osiris 2, (ventilator b), Oxylog 1000, (ventilator c), Oxylog 2000, (ventilator d), AXR1a, (ventilator e)] and three ICU ventilators which can be used for this purpose [Horus, (ventilator f), T-Bird, (ventilator g), and SV 300, (ventilator h)] were compared using a test lung regarding: 1) their capability to maintain set tidal volumes (V(T)) of 300 ml, 500 ml, and 800 ml under a normal condition A [resistance (R) 5 cmH(2)O/l/s and compliance (C) 100 ml/cmH(2)0] and two abnormal conditions B (R 20-C 30) and C (R 50-C 100); 2) trapped volume (expired V(T)relative to inspired V(T)at 0.7 s, 1 s, and 1.4 s), an estimate of the expiratory resistance of both circuit and valve; and 3) the triggering system assessed from the measurements of Delta t, Delta P for two inspiratory efforts at a PEEP of 0 cmH(2)0 and 5 cmH(2)0 in ventilators b, d, f, g, and h. Flow and airway pressure were measured with an independent physiologic recording system. 1) V(T). For ventilators a-h, the mean+/-SD changes of a set V(T)of 300 ml were -2.6+/-0.2%, -9.7+/-0.2%, 0+/-0%, -6.1+/-0.2%, 1.0+/-0.3%, -2.1+/-1.7%, 0.3+/-0%, and -1.3+/-0.1% ( P<0.001), respectively, during condition B relative to A. Similar results were obtained for a V(T)of 500 ml and 800 ml and during condition C relative to A; 2) Trapped volume. For ventilators a-h, trapped volume averaged 1+/-1%, 20+/-0%, 30+/-0.4%, 20+/-1%, 1+/-0%, 19+/-0%, 15+/-0%, and 14+/-0% at 0.7 s ( P <0.001) and 0.6+/-0%, 5+/-0%, 0.5+/-0%, 0+/-0%%, 0+/-0%, 0.6+/-0%, 0+/-0%, and 0+/-0% at 1.4 s ( P=NS); and 3) the triggering system of Oxylog 2000 was poor whereas it was of good quality for Horus, T-Bird, SV 300, and Osiris 2. The small portable ventilators presently investigated varied between each other and were less accurate than ICU ventilators.
Medical subject headings
- Critical Care
- Transportation of Patients
- Ventilators, Mechanical