The effects of low-intensity pulsed ultrasound on bioabsorbable self-reinforced poly L-lactide screws.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 12059023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effects of low-intensity pulsed ultrasound on the mechanical and molecular properties of self-reinforced poly L-lactide (SR-PLLA) screws were studied in vitro. SR-PLLA screws of 4.5 mm diameter were exposed on low-intensity ultrasound for 1, 3. 6, 9, and 12 weeks. After exposure, the bending strength, shear strength, and molecular weight were investigated. There were no differences in the investigated properties between the ultrasound exposure and control groups. We found no evidence that low-intensity ultrasound has any effect on the mechanical or molecular properties of SR-PLLA screws in vitro. The present results suggest that biodegradable SR-PLLA fixation devices are compatible with low-intensity ultrasound in vitro.
Medical subject headings
- Absorbable Implants
- Biocompatible Materials
- Bone Screws
- Lactic Acid
- Polyesters
- Polymers