The effect of total knee arthroplasty design on extensor mechanism function.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 12066269.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effect of total knee arthroplasty design on extensor mechanism function was evaluated prospectively in a consecutive, single-surgeon series. Group 1 knees (n = 83) were implanted with a prosthesis that has a multiradius femoral component, and group 2 knees (n = 101) were implanted with a femoral prosthesis that has a single flexion-extension axis. Group 2 knees gained flexion more rapidly: At 6 weeks, the average flexion for was 94.5 degrees for group 1 and 107.1 degrees for group 2 (P<.001). Significantly more patients with group 2 knees were able to arise from a 16-inch (low) chair without using their arms starting at 6 weeks, and a difference was maintained through 2 years (90% vs 73%; P=.003). Patients with group 2 knees had significantly less anterior knee pain when rising from a seated position starting at 6 weeks, and a difference was maintained through 2 years (1% vs 22%; P=.001).
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Osteoarthritis, Knee
Anatomy
- knee