Locking mechanism failure in the Harris-Galante porous acetabular component associated with recurrent hip dislocation.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12066286.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this review of longitudinal studies of the Harris-Galante porous acetabular component, 20% of all revisions were caused by failure of the polyethylene liner locking mechanism. We report 2 cases of locking mechanism failure in the Harris-Galante porous acetabular component presenting with recurrent dislocation. Broken tines from the acetabular locking mechanism were found embedded into the polyethylene liners in both cases. In a 79-year-old woman with recurrent dislocation, the polyethylene liner and femoral head were replaced, and the acetabular component was revised. In a 74-year-old man with chronic recurrent dislocation, the polyethylene liner and femoral head were replaced, but the acetabular component was preserved. Appropriate treatment for locking mechanism failure with a well-fixed acetabular component depends on many factors, and further long-term data are needed.
Medical subject headings
- Hip Dislocation
- Hip Prosthesis
- Prosthesis Failure