Oesophagotracheal perforation after intraoperative transoesphageal echocardiography in cardiac surgery.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 12066741.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although transoesophageal echocardiography (TOE) can be considered a safe procedure, severe complications may occur. We report an oesophagotracheal perforation diagnosed 7 days after a complex and very long four-valve replacement procedure in a patient with a poor preoperative condition. We believe that an ischaemic lesion of the oesophagotracheal wall caused by the TOE probe was the initial event leading to this perforation. This observation raises concerns about the safety of prolonged TOE monitoring and suggests that a combination of risk factors (i.e. a small stature, a very long procedure, congestive heart failure, and a low cardiac output before and after cardiopulmonary bypass) may warrant increased precautions while performing TOE during cardiac surgery.
Medical subject headings
- Echocardiography, Transesophageal
- Esophagus
- Heart Valve Prosthesis Implantation
- Monitoring, Intraoperative
- Trachea