A new, simple, and accurate method for non-invasive estimation of pulmonary arterial pressure.

Xu, J; Durand, L-G; Pibarot, P · Heart · 2002

prospective_cohort · Level II

Where this comes from

Abstract

To develop and validate a new non-invasive method for the estimation of pulmonary arterial pressure (PAP) based on advanced signal processing of the second heart sound. Prospective comparative study. Referral cardiology centre. This method was first tested in 16 pigs with experimentally induced pulmonary hypertension and then in 23 patients undergoing pulmonary artery catheterisation. The heart sounds were recorded at the surface of the thorax using a microphone connected to a personal computer. The splitting time interval between the aortic and the pulmonary components of the second heart sound was measured using a computer assisted spectral dechirping method and was normalised for heart rate. The systolic PAP varied between 14-73 mm Hg in pigs and between 20-70 mm Hg in patients. The normalised splitting interval was measurable in 97% of the recordings made in pigs and 91% of the recordings made in patients. There was a strong relation between the normalised splitting interval and the systolic PAP (pigs: r = 0.94, standard error of the estimate (SEE) = 5.3 mm Hg; patients: r = 0.84, SEE = 7.8 mm Hg) or the mean pulmonary pressure (pigs: r = 0.94, SEE = 4.1 mm Hg; patients: r = 0.85, SEE = 5.8 mm Hg). This study shows that this new non-invasive method based on advanced signal processing of the second heart sound provides an accurate estimation of the PAP.