Minimal 11-year follow-up of extensively porous-coated stems in femoral revision total hip arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 12068424.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Between 1984 and 1989, 188 consecutive femoral revision surgeries were performed. Eighteen patients died or were lost to follow-up, and the remaining 170 patients were followed for 11 to 16 years (mean, 14.2 years). Radiographic evidence of a bone ingrown stem was present in 82% of the hips, stable fibrous fixation was present in 14% of the hips, and 4% of the hips were unstable. Six stems were revised to a larger, fully coated cementless implant. Proximal femoral osteolysis was seen in 23% of femora but was limited to Gruen zones 1 and 7. No diaphyseal osteolysis was seen. The overall mechanical failure rate in this series was 4.1%. Failure of fixation correlated highly with extent of bone loss present at the time of surgery. On the basis of the radiographic and clinical results at a mean follow-up of 14.2 years, we recommend the use of extensively coated femoral stems in revision hip arthroplasty.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Coated Materials, Biocompatible
- Hip Prosthesis
Anatomy
- femur
- hip