Polyethylene liner cementation into fixed acetabular shells.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12068431.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A patient presenting with a secure cementless acetabular component and with femoral head penetration through the polyethylene liner is a common clinical problem. Cementing a new liner into the fixed shell is one option. We evaluated this option in a clinical series of 17 cases and with a preliminary mechanical study. In the 1 clinical failure (5.9%), the failure occurred at the cement-liner interface. The most important variable in optimizing the mechanical strength of the cemented liner construct was adequate preparation of the cement-liner interface. This approach to treating the patient with a fixed cementless shell and a worn polyethylene liner can provide a durable construct with minimal morbidity.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- hip