Effect of tear film break-up on higher-order aberrations measured with wavefront sensor.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12095817.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To investigate whether optical wavefront aberrations vary with tear film break-up. Observational case series. Higher-order aberrations were examined for 20 eyes of 20 normal subjects with a Hartmann-Shack wavefront sensor before and after tear film break-up. Higher-order aberrations for photopic vision (central 4 mm diameter) after tear film break-up increased 1.44 fold compared to higher-order aberrations before tear film break-up (P =.001, paired t-test). Higher-order aberrations after tear film break-up for scotopic vision (central 6 mm diameter) were also 1.23 times higher than those before break-up (P =.005, paired t-test). Wavefront sensing enabled us to evaluate the induced irregular astigmatism caused by tear film break-up quantitatively. Wavefront aberrations should be measured carefully to avoid the effects of tear film break-up, especially in wavefront-guided refractive surgery.
Medical subject headings
- Astigmatism
- Cornea
- Tears