Optimal frequencies for electric stimulation using medium-frequency alternating current.

Ward, Alex R; Robertson, Valma J; Makowski, Raphael J · Arch Phys Med Rehabil · 2002

rct · Level II

Where this comes from

Abstract

To determine the effect of single cycles of alternating current of various frequencies on sensory and motor thresholds and on relative thresholds (motor threshold/sensory threshold). Repeated-measures design. Laboratory setting. University student and staff volunteers (N=16; mean age, 34y). Single cycles of sinewave frequencies between 1 and 35kHz were delivered at 50Hz. The frequencies were applied in a random order. The motor and sensory thresholds were recorded at each applied frequency. Both sensory and motor thresholds showed a smooth decrease to a minimum at approximately 3kHz. The relative threshold reached a minimum close to 9kHz. Comparison with previous studies showed that although absolute thresholds reach a minimum at a frequency that depends on electrode size, the frequency at which the relative threshold is a minimum was independent of electrode size and independent of whether the stimulus was applied as single pulses or in burst mode. The optimal frequency for transcutaneous stimulation using medium-frequency alternating current depends on the outcome measure used. It would therefore be desirable for clinical stimulators to provide a selection of carrier frequencies.