Validation of the Diesse Mini-Ves erythrocyte sedimentation rate (ESR) analyzer using the Westergren ESR method in patients with systemic inflammatory conditions.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 12109848.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Diesse Mini-Ves (DMV) erythrocyte sedimentation rate (ESR) analyzer was designed to improve efficiency in determining the ESR. We compared the Westergren ESR method with the 4-sample DMV ESR analyzer for performance and clinical correlation. This prospective observational study, conducted at a 450-bed tertiary medical center, evaluated 291 paired samples from subjects with various systemic inflammatory conditions. Linear regression analysis revealed a statistically significant correlation between the 2 methods. Satisfactory precision of the DMV analyzer was obtained for high and mid-range ESR values. The 4-sample DMV ESR analyzer was precise and comparable in results to the Westergren ESR method. This DMV ESR analyzer is now used at our medical center based on quality control improvements that include a faster, safer, and more standardized ESR method. Hospital or office-based clinical laboratories should consider using the 4-sample DMV ESR analyzer in place of the Westergren method.
Medical subject headings
- Blood Sedimentation
- Hematology
- Inflammation