A roentgen stereophotogrammetric analysis of unicompartmental knee arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 12168169.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Roentgen stereophotogrammetric analysis (RSA) was used to investigate the stability of the tibial component in unicompartmental knee arthroplasty. Twenty patients were implanted with a Howmedica Duracon UNI prosthesis (Limerick, Ireland) and studied for an average follow-up of 30 months. In most cases, the displacements detected at 1-year follow-up were small (ie, not exceeding 0.5 mm for translations and not exceeding 1.0 degrees for rotations). Only 1 patient showed signs of mechanical loosening as the prosthesis continued to migrate, reaching 3.2 mm of total motion at 4-year follow-up. RSA did not show any deformation in the polyethylene except for the case of loosening. The clinical results were excellent and good according to the Hospital for Special Surgery scoring system. No correlation was found among demographic, clinical, and RSA data.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Equipment Failure Analysis
- Knee Prosthesis
- Photogrammetry
- Radiography
Anatomy
- knee
- tibia